Showing posts with label Hospital Engineering. Show all posts
Showing posts with label Hospital Engineering. Show all posts

Computed Radiography

With the advent of digital systems for capture, there has been a paradigm shift in the way images are captured, processed and finally presented

Vivek Bhargava
Ever since that fateful Sunday before Christmas of 1895, when Wilhelm Conrad Roentgen took a ’shadow graph’ of his wife Bertha’s hand with the wedding ring et al, the world of science and medicine changed forever. As early as January 1896, ‘images’ of fractured bones were being taken and till today the quest continues for better ways of looking ‘inside’ the human body to treat the patient better.
In this article, we will look at how X-ray imaging has evolved over the years, and while we will review all aspects of the X-ray imaging chain, we will focus more on the acquisition aspect and also take a look forward on newer technologies, which are emerging in the market.
X-ray generators range from the simple, single phase and three phase two pulse and 12 pulse types to the more recently offered high frequency types.
High frequency generators which offer the advantage of reduced exposure, minimised soft radiation and significantly lower skin dose are gaining in popularity and are now available virtually across the entire spectrum of X-ray systems (from the smallest portable device to the most sophisticated cath labs). Similarly, X-ray tubes range in capacity and sophistication from the smallest stationary anode types used in portable and mobile units to the so-called ‘zero heating’ high capacity tubes used in modern multi-slice CT scanners.
While one may have the best of X-ray generators and tubes, it is equally important to have an excellent capture system to ensure that the final output reflects the maximum possible information for the radiologist. In recent times, with the advent of digital systems for capture, there has been a paradigm shift in the way images are captured, processed and finally presented.
Before we look at the differences between the conventional (or analog) method of capturing images and the digital capture systems, it is important to understand that as far as the process of acquisition is concerned, there is no difference ie the X-ray generator, tube etc. remain the same in either method of capture. Hence, the difference they make is beyond the purview of the current discussion.

Block diagram of a typical CR system
Steps in CR System
Exposing the storage phosphor screen: In a CR system, the CR cassettes (which contain the storage phosphor screens) are exposed exactly like a conventional cassette (in a bucky or spot film device on the X-ray table or on a chest stand or vertical bucky). When the X-ray photons (having passed through the patient’s body) fall on the storage phosphor screen, close to 50 per cent of their energy is released in the form of fluorescence and the rest of it produces a latent image. Here comes the vital difference between conventional systems and CR systems. In a conventional system, it is the fluorescent image formed on the rare earth screen, which forms a latent image on a film, which has been loaded in the cassette in a dark room.
A digital cassette has no film inside it; the latent image, which has been formed on the phosphor screen, is the one, which will be subsequently ‘read’ to produce the final image. Typically, for up to eight hours after a latent image has been formed on a storage phosphor screen, the data can be ‘read’ from it without significant loss of information.
Stimulating the phosphor: Inside the CR system, there is a high-intensity light that stimulates the phosphor molecules and luminesces them. A laser beam (with a suitable optical system consisting of a lens and a galvanometer) is used to read the luminescent phosphor screen across its length and breadth.
Changing light energy to analog signal: Typically, photomultiplier tubes (PMTs) are used to receive the signals; these are devices that emit an electric signal in proportion to the quantum of light collected by them. Gain adjustment and calibration of the PMTs is an important part of optimising a CR system to produce optimal quality images. Converting analog to digital: This consists of an Analog to Digital Converter (ADC) and most good CR systems typically have a 12-bit output (i.e. 4,096 levels of signal).
Processing the digital signal: Since the signal is now in digital form, various processing techniques can be used to enhance it like edge enhancement, changing the brightness (level) or its contrast (window). Virtually, all reputed manufacturers now offer the final digital output or image in standard DICOM format. Every CR system thus includes a computer with a suitable monitor to allow the radiologist to view the images and process them to allow optimal presentation.
Printing the image: Depending on the needs of the institution, the images can be printed on a suitable printer (like CT or MRI images) or can be burned on a CD or DVD.
Erasure and re-use: After the laser has read the phosphor plate, the plate is exposed to high intensity light to ‘erase’ it, i.e. to bring the phosphor molecules back to an energy state where they can produce a latent image when stimulated by X-ray photons. Thus, the same screen can be used over and over again.

Multi-purpose DR system

A prototype mobile X-ray unit with integrated CR reader
DR Systems
CR system uses digital cassettes with phosphor screens. This means that for every exposure, the radiographer has to place the cassette, position the patient, shoot X-rays and then take the CR cassette back to the digitiser for obtaining the image as outlined in the steps above. When the volume of X-ray investigations is large (say for example more than 100 or 150 chest X-rays to be taken in a couple of hours), one approach is to have multiple X-ray rooms with CRs in each room or a heavy duty, multi-cassette CR system. Another more elegant and productive way is to use Direct Radiography (DR).
Some manufacturers rather erroneously expand DR as Digital Radiography and use the term interchangeably with CR systems, but for the purpose of this article the term DR refers to Direct Radiography. In a DR system, rather than using digital cassettes with phosphor plates, a ‘detector’ is used; this replaces the conventional bucky or spot film device or the chest stand or vertical bucky, as the case may be. This detector array (which is typically at least 17″ x 17″ wide to take care of all examinations) consists of either Charge Coupled Devices (CCD) or Cesium Iodide (flat panel detector).
Internally, the thallium doped cesium iodide phosphor is physically coupled to a large area amorphous silicon flat panel array, which gives the detector its name. The incident X-ray photons cause the cesium iodide layer to produce light whose intensity is measured by the photodiodes, which are formed by the silicon.Each pixel of the detector forms a Thin Film Transistor (TFT) from which the signal is read out and typically a 17″ x 17″ array consists of at least a 3,000 x 3,000 array matrix with a 14 bit output (8,192 levels).Currently, Flat Panel Detectors (FPDs) have become extremely popular and have proved their worth in the most demanding of applications like cath labs, where besides the larger frame rates involved, a good spatial resolution is desired along with stability and durability. The speed and productivity advantage are obvious.
Traditionally, most DR manufacturers integrate a good high-quality X-ray system (typically HF, 50 KW and above) to take full advantage of the excellent acquisition system; but there are also some manufacturers who offer ‘portable’ detectors that can be coupled to existing X-ray systems.
However, since the cost of a good high-quality X-ray system is not very large in comparison with the larger investment on the detection system, usually it does make good sense to go in for an integrated system that combines a good X-ray generator, multi-position arm or table and a FPD. The diagram below shows an example of a multi-purpose DR system where the same detector is used for various radiographic investigations
Pick and Choose
So when should one go in for DR rather than CR? The answer lies in the fact that these are not alternative technologies but rather complimentary solutions for digital radiography.
Typically a busy radiography department will have the DR in their main investigation room where large volumes of x-ray investigations are done and in their other x-ray rooms or out-patient / trauma / other departments could have CRs all of which can of course be networked CR cassettes can also be used in conjunction with DR systems for difficult cross-table angular examinations or on immobile patients. Again, while the CR cassettes must be processed separately, special software allows integration of CR images into the patient’s exam file so that CR and DR images can be viewed together, in much the same manner as some workstations which can display CT and MR images of the same patient at the same time on one console.
Future Trends
Compactness and ease of use will be the themes dominating the world of digital radiography and already a prototype mobile X-ray unit with integrated CR reader was recently displayed at RSNA 2005.
In DR, the use of thinner, flexible substrates and on-substrate magnification amplification and multiplexing circuitry; which can reduce the costs associated with Application Specific Integrated Circuits (ASICs) will help to bring down costs. We can look forward to portable DR detectors with automatic exposure control in-built in the flat panel array and ‘plug and play’ connectivity. As CR and DR technologies evolve, they are expected to expand into new applications, by complementing each other and offer healthcare providers better options for patient care. More power to the digital era!
The writer is General Manager-Digital Capture Medical Imaging Carestream Health
E-mail- vivek.bhargava@carestreamhealth.com

X-ray Cassettes & Screens Maintenance

Functions

  • Sensitivity of an X-Ray film to direct X-Ray photons is very low. The efficiency of X-Ray film to absorb X-Ray photons is only 1%. 99% of X-Ray photons are wasted.
  • Intensifying screen has a greater efficiency to absorb X-Ray photons and convert them into ultraviolet light or visible light.
  • Screen reduces exposure time and X-Ray dose.
  • Screen reduces kinetic unsharpness.
  • Cassette provides lightproof enclosure for an X-Ray film, which is sandwiched between a pair of screen.
Cassette provides contact between film and screen required for a sharp image by expelling out air trapped between film and screen.
Absorption, Conversion & Emission

Technology

Construction Elements: Cassettes
  • An aeronautical grade aluminum alloy is formed to give a curved profile. The curvature is optimized for displacement of air trapped inside the cassette body.
  • A lead sheet is affixed inside the cassette shell to prevent backscatter.
  • An open cell structured polyurethane foam in the tray allows the trapped air to escape and facilitates good film-screen contact.
  • The front plate of the cassette is made from an aluminum alloy which resists bending pressure, exerted on it when the cassette is closed.
  • Each construction element has a controlled tolerance of 0.1 mm and is engineered to eliminate gaps and leakages.
Construction Elements: Cassettes
Technical Parameters of Screens
Construction Elements: Screens
Types of Phosphor

Radiation Protection Apparel

Maintenance & Care

Storage
  • Store in dust-free environment.
  • Avoid exposure to extreme environmental conditions.
  • Storage area should be free of X-Ray radiation and chemical fumes.
  • Do not keep any weight on cassette or screen.
Cleaning
  • Periodical cleaning of screen & cassette ensures high image quality.
  • Moisten a cellulose cloth (non-fluffy) with antistatic cleaning agent.
  • Wipe the cleaner softly and evenly over the whole surface of the screen.
  • Remove dust from corners of cassette.
  • Leave the cassette open for approximately 10 minutes to enable the solvent to evaporate completely.
  • Close the cassette after screen surface has dried completely.
  • Never put excess cleaner on screen surface.
  • Never put the cassette on end for drying. This may lead to distortion of screen.
Mounting Screen in Cassette
  • Clean cassette to remove dust particles.
  • Ensure that the cassette is not warped or bent and that it closes perfectly.
  • If the cassette is warped or bent, replace it.
  • Peel off the protective liners of adhesive tape from the front screen.
  • Place front screen in the center in the front tray of the cassette, tape side down, and gently press along the edges.
  • Peel off the protective liners of the back screen and place it over the front Screen, tape side up, making sure that the edges of both screens coincide exactly.
  • Gently close and lock the cassette and wait for half an hour for perfect adhesion.
  • Enter installation date on the enclosed cassette label and affix it to the back of the cassette.

Standards & Testing

Standards
Numerous standards have been developed in various countries for cassettes and screens
  • These include DIN (Germany), ANSI, IEC, JS (Japan), MIL (U.S. Military), ISI (India) and others.
  • With emergence of ISO standards in Europe, various global standards are veering towards ISO standard 4090.
  • The ISO 4090 and ISO 4090-1 standards for cassette and screen define dimension, dimensional tolerances, geometrical accuracy, absorption standards, design parameters as well as stipulate testing methods.
Testing
Key areas for performance measurement of cassettes and screen, on the basis of ISO 4090 are:
Film-Screen Contact Test
  • Contact is the key measurement of radiological image quality of the cassette.
  • Good contact leads to accurate diagnosis.
  • Contact test should be ideally done once a year.

Film-Screen Contact Test

Film-screen contact has a significant influence on radiographic image quality. Film-screen contact leads to accurate diagnosis. The cassette should be tested for contact at least once in a year.
Procedure
  • Clean the screen and cassette with screen cleaning solution recommended by manufacturer.
  • Allow screen to dry completely.
  • Load cassette with film and wait for approximately 3 minutes for trapped air to escape out of the cassette.
  • Place a wire grid (3.15 mm mesh width for conventional radiography) on top of the cassette.
  • Expose the wire grid at 70 kV, so that exposed film has density of approx. 2.8.
  • Process the film and view an illuminator from a distance of at least 1.5 meters.
Observation
  • Appearance of overall uniform density indicates uniform film-screen contact.
  • Areas, which appear dark, indicate areas of poor or lack of contact. Such dark areas in diagnostically important parts of the image are not acceptable.
  • Please view good contact and poor contact images.
Good Contact
Good Screen Contact
Poor Contact
Poor Screen Contact
Speed & Resolution Test
  • Quantum of X-Ray dosage required to raise the density of the X-Ray film to the desired level depends on the Screen speed.
  • A screen speed of 400 together with the use of a regular speed film addresses the requirements of most of the general radiological examinations.
  • Resolution of screen is the key to clear film reading.

Speed & Resolution Test

Procedure
  • Load the pair of test screen and reference screen in the standard cassette.
  • Load the X-Ray film.
  • Use pre-exposed step wedge as an object.
  • Expose step-wedge at 110 FFD, 70 kV, 8 mAs.
  • Develop the exposed film in automatic processor. Measure density by using densitometer.
  • Measure density for 3 steps of the step-wedge for test screen and reference screen exposed. Record the readings.
  • Density to measure at marked center step and step above and below it for 3 step readings. The density of reference screen center step should be between 1.2 to 1.5 D. Adjust mAs to get correct density.
  • Plot a graph of Density vs. Log IT (I is current & T is time), considering that each step corresponds to 0.05 IT. Use a scale of 15 mm = 0.05 IT on X-axis, and suitable scale on Y-axis as per density values.
  • Draw a line perpendicular to X-axis at density value corresponding to center step for the reference screen.
  • Draw a line perpendicular to X-axis from point of intersection of line drawn from center point of reference screen and line graph of test screen.
  • Calculate difference of log IT on X-axis. It is the difference of the reading of test screen and reference screen.
Calculate speed as shown below:
A = (Difference x 0.05)/15
B = Log constant for reference screen
Speed = Antilog of B x 100.
Measurement of Resolution by X-Ray Exposure
  • Use same pair of screen for resolution test as used for speed testing.
  • Expose test screen & reference screen with resolution test kit as an object.
  • Develop the exposed film in automatic processor. Measure density of exposed film at area 1 inch below the resolution image. The density should be between 2.2 to 2.6 D.
  • Measure resolution by viewing the film illuminator using 10x lens. Lines clearly visible in maximum resolution step is considered as resolution and recorded for both reference & test screens.
Cassette Light-Proofness Test
  • Light leakage in diagnostically important areas, on the edges of a film leads to loss in portions of an X-Ray image.
  • The cassette has to be appropriately designed, engineered and must use optimum construction elements to result in light-proofness.

Cassette Light-proofness Test

Light-proofness of cassette is a basic requirement to protect from loss of data of a radiological examination.
Procedure
  • Cassette is fitted with a pair of screens and loaded with X-Ray film.
  • It is exposed for 10 minutes to a 100 watt frosted tungsten filament lamp located at a distance of 1 metre (39 inches) from the cassette.
Evaluation
  • Increase of density of more than 0.1 on the edge indicates light leakage.
  • Diagnostically important areas for light-proofness are all 4 edges to a maximum extent of 3 mm (1/8th inch)
Acceptable Light Proofness

Non-Acceptable Light Proofness

Cassette Durability Test
  • Durability in cassette is essential for maximizing the return on investment in cassettes.
  • A durable cassette should be able to withstand periodical accidental drops during usage.

Cassette Durability Test

Durability of a cassette is tested by carrying out a drop test.
Procedure
  • Carry out a film-screen contact test and a light-proofness test on the cassette to be tested for durability.
  • Drop the cassette from a height of 1 metre (39 inches) on a concrete floor.
  • Cassette should be dropped in such a way that it strikes each corner and each edge, for a total of 8 drops.
  • A film-screen contact test and a light-proofness test should be done after each of the 8 drops.
Evaluation
  • In case of a durable cassette, there is no deterioration in the extent of film-screen contact and light-proofness.

Copyright © Kiran Medical Systems 2003-2008.

Ways of Maximizing X-ray Tube's Life Span

Minimise filament boost (“prep”) time
Boost time will usually exceed the actual exposure time. High filament current
applied for too long will shorten filament life and will lead to unstable operation asevaporated tungsten from the filament is deposited onto the glass envelope. This is
especially the case at high mA stations.

Use lower tube current (mA)
The high filament current required to produce high tube current (mA) will shorten
filament life and will lead to unstable operation as evaporated tungsten from the
filament is deposited on to the glass envelope. Whenever possible, use a lower mA
station and a longer exposure time to arrive at the desired mAs.

Follow rating charts and anode heating/cooling curves
Operation beyond published ratings will result in premature focal track wear or
damage. Even moderate etching of the focal track will result in a fall-off in radiation
output, because electrons from the filament which strike in micro-crevices in the
target material produce radiation that is mostly absorbed in the surrounding target
material. More severe etching, or melting, results in the liberation of gasses from the
target material, which causes tube instability. Excessive heat transfer from the target
into the rotor body will cause bearing failure or slow rotation which will result in
melts on the focal track.

Limit operation to 80% of maximum single exposure ratings
although higher power levels are both possible and permitted, this reduction will help
assure long focal track life. Also, it will minimise the reduction in radiation output
associated with a roughened focal track.

Do not exceed anode thermal capacity or dissipation rate of the target
the greatest danger is to heat flow into the bearing structure, as discussed above. In addition, gasses may be emitted from the various metals within the tube if the
temperature reached during clinical use is appreciably higher than that used during the “outgassing” stage of manufacture. If outgassing occurs during clinical use, the tube will become unstable. (“Aging” the tube may reverse the process but this is not
assured.)

Do not make high mA exposures on a cold target
Uneven expansion caused by thermal stress from a high power exposure can result in
a cracked target. Do not assume that a “thermally relieved” target design provides
absolute protection. Always follow the recommended warm- up procedure. The procedure may need to be repeated between patients, if the “idle” time is long enough, in addition to being performed at the beginning of the workday.

Avoid long intervals between spot-films
Most systems provide for a “holdover” period of up to approximately 25 seconds
between spot- films, during which the rotor is kept at high speed before the rotor
brake/reboost cycle is allowed to occur. In some systems, the filament current remains at the exposure value during this period, thereby causing evaporation of filament material and resultant tube instability.

Limit rotor start/stop operations
Rotor start/stop operations especially to/from high speed (150/180 Hz) generate
considerable heat in the stator windings, which will lead to stator damage in extreme
cases. Generally there should be a minimum of 30 to 40 seconds between starts.
Tubes equipped with a heat exchanger will be less sensitive to this potential problem
because oil circulation will help prevent hot spots from occurring around the stator
windings.